Hemolysis research of implantable axial flow pump for two -step heart transplantation in children
https://doi.org/10.15825/1995-1191-2017-1-22-27
Abstract
Introduction. One of the main indicators characterizing mechanical circulatory support devices (artificial valve, implantable pumps, etc.) is trauma of blood cells. Therefore, while developing new pumps, one of the key studies in vitro is to evaluate blood hemolysis. For an objective hemolysis analysis of pump it is required to create a standardized methodology of hemolysis studies. The object of the study in this paper is implantable axial pump DON for two-step heart transplantation in children.
The aim of study is to develop a standardized methodology of hemolysis studies of blood pumps and to conduct research of pediatric axial pump DON.
Materials and methods. To conduct hemolysis research we created a mock circulatory system consisting of a reservoir placed in water bath maintaining a constant working fluid (blood) temperature, hydrodynamic resistance, connecting tubes, ports for blood sampling and pressure and flow measurement systems, and research pump. Test method is to estimate levels of free hemoglobin pHb obtained by blood samples during pump working in operating mode (for pediatric pump: blood flow 2.5 l/min, pressure difference 80 mmHg). Using the data obtained the standardized indices of hemolysis NIH and MIH are calculated based on pHb values, hematocrit, total hemoglobin, blood flow and working pump time.
Results. We developed and realized a standardized methodology of hemolysis research by which we evaluated hemolysis of pediatric axial pump. The results of hemolysis tests allowed us to optimize the design of DON. Obtained values of hemolysis of the latest version of pediatric pump DON-3 have shown that they do conform to the requirements of minimum blood injury and it allows us to proceed to the next step of pediatric pump research – animal experiments.
Conclusion. Developed methods and evaluation tools of hemolysis allow us to provide objective information on one of the most important indicators of developing implantable pediatric axial pump and they could be recommended for hemolysis research of others pumps.
About the Authors
O. Yu. DmitrievaRussian Federation
Moscow, Russian Federation
A. S. Buchnev
Russian Federation
Moscow, Russian Federation
A. A. Drobyshev
Russian Federation
Moscow, Russian Federation
G. P. Itkin
Russian Federation
Address: 1, Shchukinskaya st., Moscow, 123182, Russian Federation. Tel. (499) 190-60-34
References
1. Throckmorton AL, Chopski SG. Pediatric Circulatory Support: Current Strategies and Future Directions. Biventricular and Univentricular Mechanical Assistance. ASAIO Journal. 2008; 54: 491–497.
2. Rosenthal DN, Almond CS, Jaquiss RD, Peyton CE, Auerbach SR et al. Adverse events in children implanted with ventricular assist devices in the United States: Data from the pediatric interagency registry for mechanical circulatory support (PediMACS). J. Heart Lung Transplant. 2016; 35 (5): 569–577.
3. Zafar F, Jefferies JL, Tjossem CJ, Bryant R, Jaquiss RD, Wearden PD et al. Biventricular Berlin Heart EXCOR pediatric use across the United States. Ann. Thorac. Surg. 2015; 99 (4): 1328–1334.
4. Potapov EV, Hetzer R. Pediatric Berlin Heart Excor. Ann. Thorac. Cardiovasc. Surg. 2006; 12 (2): 15.
5. Baldwin JT, Duncan BW. Ventricular assist devices for children. Progress in Pediatric Cardiolog. 2006; 21: 173–184.
6. Fraser CD, Jaquiss RDB, MD, Rosenthal DN et al. Prospective trial of a pediatric ventricle assist device. New England J. of Medicine. 2012; 367 (6): P532–541.
7. Itkin GP, Matveev YG, Romanov OV. Comparative hemolysis tests of rotary blood pump. Artificial Organs. 1995; 19 (7): 616–619.
8. Wurzinger LJ, Opitz R, Eckstein H. Mechanical blood trauma: an overview. Angeiologie. 1986; 38: Р81–97.
9. Leverett LB, Hellums JD, Alfrey CP, Lynch EC. Red blood cell damage by shear stress. Biophys. J. 1972; 12: 257–272.
10. Throckmorton AL, Untaroiu A, Allaire PE, Houston G, Wood HG et al. Computational analysis of an axial flow pediatric ventricular assist. Artificial Organs. 2007; 28 (10): 881–891.
11. Paul R, Schügner F, Reul H, Rau G. Recent findings on flow induced blood damage: critical shear stresses and exposure times obtained with a high shear stress Couette system. Artif. Organs. 1999; 23: 680.
12. Yeleswarapu KK, Antaki JF, Kameneva MV, Rajagopal KR. A mathematical model for shear induced hemolysis. Artif. Organs. 1995; 19: 576–582.
13. Apel J, Paul R, Klaus S, Siess Th, Reul H. Assessment of hemolysis related quantitie in a microaxial blood pump by computational fluid dynamics. Artif. Organs. 2001; 25: 341–347.
14. Huang CR, Fabisiak W. A rheological equation characterizing both the time dependent and steady state viscosity of human blood. AIChE Symp. Series. 1978; whole: 19–21.
15. Affeld K, Goubergrits L, Holberndt O. Novel cardiac assist valve with a purge flow in the valve sinus. ASAIO J. 1998; 44: M642–647.
16. ASTM F1841-97: Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps. 2005.
17. Kawahito K, Nosé Y. Hemolysis in different centrifugal pumps. Artif. Organs. 1997; 21: 323–326.
18. Nakazawa T, Takami Y, Benkowski R, Ohtsubo S, Yukio O et al. Development and Initial Testing of a Permanently Implantable Centrifugal Pump. Artificial Organs. 1997; 17: 597–601.
19. Araki K, Anai T, Oshikawa M, Nakamura K, Onitsuka T. In vitro Performance of a Centrifugal, a Mixed Flow, and an Axial Flow Blood Pump. Artificial Organs. 1998; 22 (5): 366–370.
20. Kawahito K, Nose Y. Hemolysis in Different Centrifugal Pumps. Artificial Organs. 1997; 21 (4): 323–326.
21. Kobayashi K, Nitta S, Yambe T, Sonobe T, Naganuma S et al. Hemolysis test of disposable type vibrating flow pump. Artificial Organs. 1997; 21 (7): 691–693.
22. Gobel C, Eilers R, Reul H, Schwindke P, Jorger M, Rau G. A New Blood Pump for Cardiopulmonary Bypass: The HiFlow centrifugal pump. Arfificid Orgarns. 1997; 21 (7): 841–845.
23. Mueller MR, Schima H, Engelhardt H et al. In vitro hematological testing of rotary blood pumps; remarks on standardization and data interpretation. Artif. Organs. 1993; 17: 103–110.
24. Schima H, Müller MR, Tsangaris S et al. Mechanical blood traumatization by tubing and throttles in in vitro pump tests: experimental results and implications for hemolysis theory. Artif. Organs. 1993; 17: 164–170.
25. Tamari Y, Lee-Sensiba K, Leonard EF, Parnell V, Tortolani AJ. The effects of pressure and flow on hemolysis caused by Bio-Medicus centrifugal pumps and roller pumps. Guidelines for choosing a blood pump. J. Thorac. Cardiovasc. Surg. 1993; 106: 997–1007.
26. Naito K, Suenaga E, Cao Z-L, Suda H, Ueno T et al. Comparative hemolysis study of clinically vailable centrifugal pumps. Artif. Organ. 1996: 20 (6): 560–563.
27. Maruyama O, Yamaguch K, Nishida M, Onoguchi T, Tsutsui T. Hemolytic evaluation using polyurethane microcapsule suspensions in circulatory support devices: normalized index of hemolysis comparisons of commercial centrifugal blood pumps. Artificial Organs. 2007; 32 (2): 146–156.
Review
For citations:
Dmitrieva O.Yu., Buchnev A.S., Drobyshev A.A., Itkin G.P. Hemolysis research of implantable axial flow pump for two -step heart transplantation in children. Russian Journal of Transplantology and Artificial Organs. 2017;19(1):22-27. (In Russ.) https://doi.org/10.15825/1995-1191-2017-1-22-27