Hemolysis testing of a small-sized axial flow pump for implantation in patients with low anthropometric parameters

https://doi.org/10.15825/1995-1191-2026-2-76-82

Abstract

The Biotechnical Systems Laboratory assessed the trauma potential of a newly developed pediatric implantable axial flow pump. A mathematical evaluation was performed to assess shear stress, blood exposure time, and the hemolysis index. Regions contributing to blood trauma were identified at flow rates ranging from 1.5 to 2.5 L/min. A standardized methodology was developed and applied to evaluate blood cell damage (hemolysis) under simulated pediatric hemodynamic conditions, with a pressure gradient of 85 ± 5 mm Hg and three flow rates of 1.5, 2.0, and 2.5 L/min. The operation of the compact implantable axial pump prototype under left ventricular assist conditions was designed to account for the anthropometric characteristics of patients with a body surface area of less than 1.2 m2. Hemolysis tests were performed using a hydrodynamic test bench that ensures continuous circulation of donor blood. The testing protocol involved measuring the concentration of free hemoglobin in plasma samples collected during pump operation on the test bench. Based on these measurements, hemolysis and modified hemolysis indices were calculated, demonstrating positive performance dynamics of the pump under the tested conditions.

About the Authors

A. S. Buchnev
Shumakov National Medical Research Center of Transplantology and Artificial Organs
Russian Federation

Alexander S. Buchnev.

1, Shchukinskaya str., Moscow, 123182

Phone: (926) 470-09-88



A. P. Kuleshov
Shumakov National Medical Research Center of Transplantology and Artificial Organs
Russian Federation

Moscow



N. V. Vasiliev
Shumakov National Medical Research Center of Transplantology and Artificial Organs
Russian Federation

Moscow



V. A. Elenkin
Shumakov National Medical Research Center of Transplantology and Artificial Organs
Russian Federation

Moscow



N. V. Grudinin
Shumakov National Medical Research Center of Transplantology and Artificial Organs
Russian Federation

Moscow



References

1. Baldwin JT, Adachi I, Teal J, Almond CA, Jaquiss RD, Massicotte MP et al. Closing in on the PumpKIN Trial of the Jarvik 2015 Ventricular Assist Device. Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu. 2017 Jan; 20: 9–15.

2. Kilic A, Nolan TD, Li T, Yankey GK, Prastein DJ, Cheng G et al. Early in vivo experience with the pediatric Jarvik 2000 heart. ASAIO J. 2007 May-Jun; 53 (3): 374–378.

3. Roberts N, Chandrasekaran U, Das S, Qi Z, Corbett S. Hemolysis associated with Impella heart pump positioning: In vitro hemolysis testing and computational fluid dynamics modeling. Int J Artif Organs. 2020 Mar 4: 391398820909843.

4. Sobieski MA, Giridharan GA, Ising M, Koenig SC, Slaughter MS. Blood trauma testing of CentriMag and RotaFlow centrifugal flow devices: a pilot study. Artif Organs. 2012 Aug; 36 (8): 677–682.

5. Nikfar M, Razizadeh M, Zhang J, Paul R, Wu ZJ, Liu Y. Prediction of mechanical hemolysis in medical devices via a Lagrangian strain-based multiscale model. Artif Organs. 2020 Aug; 44 (8): E348–E368.

6. Rowlands GW, Pagani FD, Antaki JF. Classification of the Frequency, Severity, and Propagation of Thrombi in the HeartMate II Left Ventricular Assist Device. ASA-IO J. 2020 Sep/Oct; 66 (9): 992–999.

7. Dai WF, Wu P, Liu GM. A two-phase flow approach for modeling blood stasis and estimating the thrombosis potential of a ventricular assist device. Int J Artif Organs. 2021 Jul; 44 (7): 471–480.

8. Dang G, Epperla N, Muppidi V, Sahr N, Pan A, Simpson P, Baumann Kreuziger L. Medical Management of Pump Related Thrombosis in Patients with Continuous Flow Left Ventricular Assist Devices: A Systematic Review and Meta-analysis. ASAIO J. 2017 Jul/Aug; 63 (4): 373–385.

9. Buchnev AS, Kuleshov AP, Drobyshev AA, Elenkin VA, Shevchenko GA, Grudinin NV. Bench studies of a small axial pump for implantation in patients with low anthropometry. Russian Journal of Transplantology and Artificial Organs. 2025; 27 (2): 81–88. https://doi.org/10.15825/1995-1191-2025-2-81-88.

10. Itkin GP, Shemakin SJ, Shokhina EG, Burcev VI, Avramov PV, Volkova EA et al. The first domestic implantable axial flow pump: results of experimental studies in calves. Russian Journal of Transplantology and Artificial Organs. 2013; 15 (3): 49–58. (In Russ.). https://doi.org/10.15825/1995-1191-2013-3-49-58.

11. ASTM F1841-97. Standard Practice for Selection of Blood for In Vitro Evaluation of Blood Pumps. West Conshohocken, PA: American Society for Testing and Materials, 2017.

12. ASTM F1841-25. Standard Practice for Selection of Blood for In Vitro Evaluation of Blood Pumps. West Conshohocken, PA: American Society for Testing and Materials, 2025.

13. Nosé Y. Design and development strategy for the rotary blood pump. Artif Organs. 1998 Jun; 22 (6): 438–446.

14. INTERMACS Adverse Event Definitions, Appendix A: Adverse Event Definitions Version 2021.

15. ASTM F1840-97. Standard Practice for Selection of Blood for In Vitro Evaluation of Blood Pumps. West Conshohocken, PA: American Society for Testing and Materials, 2013.


Review

For citations:


Buchnev A.S., Kuleshov A.P., Vasiliev N.V., Elenkin V.A., Grudinin N.V. Hemolysis testing of a small-sized axial flow pump for implantation in patients with low anthropometric parameters. Russian Journal of Transplantology and Artificial Organs. 2026;28(2):76-82. (In Russ.) https://doi.org/10.15825/1995-1191-2026-2-76-82

Views: 33

JATS XML


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1995-1191 (Print)